Specifications - BD Alaris 8002TIG03 Directions For Use Manual

Syringe pump with plus software mk4
Hide thumbs Also See for Alaris 8002TIG03:
Table of Contents

Advertisement

Specifications

Infusion Specifications
Maximum infusion rate can be set as part of the configuration .
0 .1ml/h - 150ml/h
5ml syringes
0 .1ml/h - 300ml/h
10ml syringes
0 .1ml/h - 600ml/h
20ml syringes
0 .1ml/h - 900ml/h
30ml syringes
0 .1ml/h - 1200ml/h
50ml syringes
Infusion Rate Increments:
Rate Range (ml/h)
Single Chevron Key Increments (ml/h) Double Chevron Key Increments (ml/h)
0 .10 to 9 .99
0 .01
10 .0 to 99 .9
0 .1
100 to 999
1
1000 to 1200
10
The Volume Infused range is 0 .0ml - 9990ml .
Bolus Specifications
Maximum Bolus rates can be set as part of the configuration . Bolus rates are user adjustable, in increments of 10ml/h .
10 ml/h - 150ml/h
5ml syringes
10 ml/h - 300ml/h
10ml syringes
10 ml/h - 600ml/h
20ml syringes
10 ml/h - 900ml/h
30ml syringes
10 ml/h - 1200ml/h
50ml syringes
The bolus volume limit can be set as part of the configuration .
• Minimum: 0 .1ml; maximum 25 .0ml
• Increments of 0 .1ml; default 5 .0ml
During BOLUS the pressure limit alarms are temporarily increased to their maximum level .
Bolus Volume Accuracy*
Bolus Volume
0 .1ml
25ml
Bolus Volume
GH
0 .1ml
25ml
* - Using BD Plastipak 50ml syringe at 5ml/h under normal conditions (95% confidence / 95% of pumps) .
Critical Volume
The bolus which can occur in the event of a single internal fault condition with a 50 ml syringe is : Maximum Overinfusion - 0 .87ml
Purge Specifications
The purge rate is limited to the maximum rate for the syringe and can be set as part of the configuration .
100ml/h - 500ml/h .
The purge volume range is 0 .5ml - 5ml .
During PURGE the pressure limit alarms are temporarily increased to their maximum level .
Keep Vein Open (KVO) Rate
0 .1 ml/h - 2 .5ml/h .
End Of Syringe Rate
Stop, KVO (0 .1ml/h to 2 .5ml/h), or set rate if lower than KVO .
Volume To Be Infused (VTBI)
0 .10ml - 1000ml, 1min - 24h
Typical
Typical Maximum Typical Minimum
1 .7%,
5 .1%
0 .1%
0 .5%
Typical
Typical Maximum Typical Minimum
1 .9%
6 .2%
0 .2%
0 .5%
BDDF00001 Issue 5
0 .10
1 .0
10
100
Pump Specification
-2 .5%
± 10%
-0 .6%
± 5%
Pump Specification
-7 .3%
± 10%
-0 .1%
± 5%
37/56
Alaris™ Syringe Pumps
Specifications

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Alaris 8003tig03-gAlaris 8003tig03Alaris 8002tig03-g

Table of Contents