Display
TITRATION NOT
CONFIRMED
VTBI DONE
(KVO/CONTINUE)
AC POWER FAIL
NEAR END OF INFUSION
ADD DRUG NOT COMPLETE
ATTENTION
*Configurable option .
Note: The audio sound pressure level is at least 45 dB depending on configuration of the alarm sound level .
Setting the alarm sound pressure level lower than the ambient sound pressure level can impede User recognition of
w
alarm conditions.
Alarm Priority Level Indicators
Priority
Audio Indicator
HIGH
One urgent tone pulse followed by one second pause
MEDIUM
One warning tone pulse followed by one second pause
LOW
Three attention tone pulse followed by a three second pause
Alarm
Description and Troubleshooting Guide
Priority
Medium
The infusion rate has been changed, but has not been confirmed after five seconds of no
activity the User will be notified by an auditory cue . The infusion has not been confirmed
and two minutes has expired without any operation, a medium priority alarm will be
generated . Press the
clear this message and silence the alarm . Check infusion rate and confirm by pressing the
b
button or press the
rate titration is enabled) . Pressing QUIT will cancel the titration and keep the original rate .
Medium
The pre-set Volume To Be Infused is complete and the Pump continues to infuse at set rate
or at KVO rate .
Medium
AC Power has been disconnected and the Pump is operating on battery power, if this
occurs when the Pump is infusing the message INFUSION CONTINUES will be displayed .
Reconnect AC power supply or press the
with battery operation . The alarm will automatically cancel if the AC power supply is
reconnected .
Medium
The Pump is nearing the end of the infusion . This value can be configured . Optional*
reminder signals may sound, they are auditory signals that have four beeps that occur
every ten minutes once the NEOI alarm is cancelled . This reminder signal will not occur for
a continuous infusion in which the NEOI alarm is set for less than ten minutes .
Medium
The ADD DRUG operation has not been completed after five seconds the User will be
notified by an auditory cue . After two minutes have expired and the ADD DRUG operation
has not been completed, a low priority alarm is generated . Press CANCEL softkey and
complete ADD DRUG operation .
If the Pump has been left on for more than 2 minutes* (referred to as ATTENTION in the
Low
log) without starting the operation, a low priority alarm will be generated . Press the
button to silence the alarm for a further 2 minutes . For extended Attention Timeout press
and hold down the
Pump on standby for 15 minutes .
BDDF00001 Issue 5
R
button to silence the alarm, then press the CANCEL softkey to
h
button to revert to the previous rate . (This alarm only occurs if
R
button to silence the alarm and continue
R
button and wait for four beeps in succession, this will put the
28/56
Alaris™ Syringe Pumps
Alarms and Warnings
R
Visual Indicator (Beacon)
Flashing Red
Flashing Amber
Flashing Amber