Dimensions
335 mm (w) x 121 mm (h) x 200 mm (d) .
GH
310 mm (w) x 121 mm (h) x 200 mm (d) .
Weight
2 .4 kg (excluding power cable) .
Protection against fluid ingress
IP32 - Protected against direct sprays of water up to 15° from vertical and protected against solid objects greater than 2 .5mm .
Note: IP33 applies if mains retainer kit, part number 1000SP01294, is fitted .
Alarm Conditions
Drive Disengaged
Check Syringe
Line Occlusion
Near End Of Infusion
Internal Malfunction
Dose Would Exceed
Bolus Dose Under
Concentration not Permitted
Environmental Specifications
Operating Temperature
Operating Relative Humidity
Operating Atmospheric Pressure
Transport & Storage Temperature
Transport & Storage Relative Humidity
Transport & Storage Atmospheric Pressure
Electrical/Mechanical Safety
Complies with EN/IEC60601-1 and EN/IEC60601-2-24 .
Potential Equalisation Conductor
The function of the Potential Equalisation Connector (Conductor) is to provide a direct connection between the Pump and the potential
equalisation busbar of the electrical installation . To use the Potential Equalisation Connector, connect the Potential Equalisation
Connector on the Pump to the potential equalisation busbar of the electrical installation .
EMC
Complies with EN/IEC60601-1-2 and EN/IEC60601-2-24 .
Occlusion
Battery Low
Battery Empty
End of Infusion
Pressure Disc Out
Dose not Permitted
Bolus Dose Over
Weight Outside Limit
0°C - +40°C
20% - 90%
700hPa - 1060hPa
-30°C - +50°C
10% - 95%
500hPa - 1060hPa
BDDF00001 Issue 5
Attention (Nurse Callback)
Titration not confirmed
VTBI Done
AC Power Fail
Dose Under
Bolus Dose not Permitted
Rate not Permitted
Add Drug Not Complete
39/56
Alaris™ Syringe Pumps
Specifications