Securing the Syringe with Optional Lock box

Lock box operation

The optional lock box is a available in two configurations:
• Rate Unlocked lock box - is designed to allow the user to
adjust rate whilst infusing .
When mounting a Pump with lock box ensure that there is sufficient clearance for the cover to be fully opened, a gap
w
of 130mm minimum below the Pump is recommended.
Open Lock box:
1 . Insert the key into the lock and turn key either way to unlock .
Close Lock box:
1 . Load the syringe according to the instructions in the 'Loading and Confirming a Syringe' section .
2 . Ensure the extension set is connected to the syringe and threaded through the lock box .
Note: For the Alaris CC Syringe Pump models the pressure disc must be fitted prior to closing the lock box .
3 . Set up the Pump according to the instructions in the 'Starting the Pump' section prior to closing the lock box for the Rate Locked
version only .
4 . Close the cover until it makes contact with the Pump case .
6 . Remove the key .
• Pumps with lock box fitted should only be used when fitted to an Alaris Gateway Workstation, Alaris DS Docking
w
Station or an I.V. Pole.
• When transporting the Pump with lock box fitted it is recommended that two hands are used when holding or
carrying the Pump.
• If lock or lock box appears to be damaged, remove the Pump from service for examination by Qualified Service
Personnel.
• When the Pump is not in use ensure the lock box is locked.
• Lock box keys should be stored separately and take care not to lock keys inside lock box.
• Refer to the 'Routine Maintenance' section for instructions on cleaning and maintenance.
BDDF00001 Issue 5
Securing the Syringe with Optional Lock box
• Rate Locked lock box - is designed to prevent rate change
whilst infusing . If using this lock box users would need to
put the Pump on hold and open the lock box to change the
rate .
2 . Lock box will move to the right and can then be opened .
5 . Push the lock box from right to left until a click is heard .