Pressure Features

Auto Set Pressure (If enabled)*

If the Auto Set Pressure Option is enabled then the Pump automatically adjusts the pressure occlusion limit .
After 15 minutes of infusion the Pump automatically adjusts the occlusion pressure limit to XX mmHg above the average infusion
pressure, taken from the average of the last five minutes of infusion .
Note: XX is the AUTO OFFSET pressure and is determined by the user . This adjustment, AUTO OFFSET value 15-100mmHg, is
configurable by profile within the data set . At pressures up to 100mmHg the AUTO OFFSET value is added . For pressures
above 100mmHg the alarm level is set to whatever the AUTO OFFSET value is as a percentage above the average infusion
pressure up to the maximum pressure defined within the data set .
*

Pressure Level with pressure set fitted*

1 . To check and adjust the pressure level press the
displaying the pressure alarm level and the current pressure level .
f
2 . Press the
keys to increase or decrease the pressure alarm level . The new level will be indicated on the display .
3 . The AUTO Pressure feature may be used when a stable pressure has been achieved over a short period of infusion . If AUTO Pressure
has been enabled the automatic pressure alarm level is calculated and set by pressing the AUTO softkey .
4 . Press the TREND softkey to view the pressure trend of the previous 12 hours . The pressure trend can be viewed at 15 minute
intervals by using the +/- softkeys . The pressure trend graph displays the pressure at a given time .
5 . Press the OK softkey to exit the pressure screen .
*

Pressure Level*

1 . To check and adjust the pressure level press the
current pressure level .
f
2 . Press the
keys to increase or decrease the alarm level . The new level will be indicated on the display .
3 . Press OK to exit the screen .
The interpretation of the pressure readings and occlusion alarms are the responsibility of the clinician and should
w
include the clinical context in which the Pump is being used.
*
without pressure set fitted (not applicable when Fully Dedicated)
e
button . The display will change to show a 20 minute pressure trend graph
20mmHg
e
button . A bar graph will be displayed showing the pressure alarm level and the
BDDF00001 Issue 5
07:45
_
+
OK
26/56
Alaris™ Syringe Pumps
Pressure Features