Dosing Summary

To review currently selected dosing information:
d
1 . Press the
button to first access the options menu .
2 . Select DOSING SUMMARY .
3 . Review the information and then press the QUIT softkey .

Set VTBI over Time

This option allows a VTBI and delivery time to be specified . The rate necessary to deliver the required volume within the specified time is
calculated and displayed .
1 . Stop the infusion . Press the
2 . Select the SET VTBI OVER TIME option using the
3 . Adjust the volume to be infused using the
4 . Enter the time over which the volume is to be infused . The infusion rate will automatically be calculated . Press the OK softkey to
enter the value .
5 . Select the rate at VTBI end from the list using the

24 Hour Log

This option allows the 24 hour log of volume infused to be reviewed .
d
1 . Press the
button to access the options menu .
2 . Select the 24H LOG option using the
The display shows the hourly volume infused . The volume infused shown in brackets is the total volume infused since the volume was
last cleared . See example below:
07:48 - 08:00 4 .34ml (4 .34ml)
08:00 - 09:00 2 .10ml (6 .44ml)
09:00 - 10:00 2 .10ml (8 .54ml)
VOLUME CLEARED
3 . Press the QUIT softkey to exit the log .

Event Log

This option allows the event log to be reviewed, if enabled .
d
1 . Press the
button to access the options menu .
2 . Select the EVENT LOG option using the
3 . Scroll through the log using the
Note: When the event log reaches full capacity, the oldest events will be overwritten by the most recent events .

Data Set Details

To review currently selected data set information:
d
1 . Press the
button to access the options menu .
2 . Select DATA SET DETAILS .
3 . Review the information and then press the QUIT softkey .

Infusion Setup

To change Infusion Setup
1 . Stop the infusion . Press the
2 . Select INFUSION SETUP .
3 . Select Infusion Setup required and press the OK softkey .

Pump Details

To review Pump information .
d
1 . Press the
button to access the options menu .
2 . Select PUMP DETAILS .
3 . Review the information and then press the QUIT softkey .
Note: The following information will be displayed:
• UNIT REFERENCE
• SN
• S/W
d
button to access the options menu .
f
f
f
f
keys and press the OK softkey .
f
keys and press the OK softkey .
f
keys . Press the QUIT softkey to exit the log .
d
button to access the options menu .
An identifier configured in Technician Mode by qualified technical personnel .
The Serial Number of the Pump
Software version of the Pump
BDDF00001 Issue 5
keys and press the OK softkey .
keys . When the desired volume has been reached press the OK softkey .
keys and press the OK softkey . The default is STOP .
24/56
Alaris™ Syringe Pumps
Basic Features