Prompts

Prompts are indicated by an audible alarm and message, they cannot be silenced and do not have a visual indicator .
Display
DOSE WOULD EXCEED*
DOSE UNDER*
DOSE NOT PERMITTED
BOLUS DOSE OVER*
BOLUS DOSE UNDER*
BOLUS DOSE NOT PERMITTED
CONCENTRATION NOT PERMITTED
WEIGHT OUTSIDE LIMIT*
RATE NOT PERMITTED
Guardrails
*
Icon
Description and Troubleshooting Guide
The dose rate has been set to a value which exceeds a Guardrails Soft Alert . Check
infusion setting, to continue with infusion at set rate confirm OVERRIDE LIMIT by
pressing the YES softkey . If OVERRIDE LIMIT is not required press the NO softkey and
adjust rate below Guardrails Soft Alert .
The dose rate has been set to a value which is under a Guardrails Soft Alert . Check
infusion setting, to continue with infusion at set rate confirm OVERRIDE LIMIT by
pressing the YES softkey . If OVERRIDE LIMIT is not required press the NO softkey and
adjust rate above Guardrails Soft Alert .
The dose rate has been set above a Hard Limit . Check infusion setting and adjust rate
to appropriate required rate .
The bolus dose has been set to a value which exceeds a Guardrails Soft Alert . Check
the bolus setting, to continue with the bolus confirm OVERRIDE LIMIT by pressing the
YES softkey . If OVERRIDE LIMIT is not required press the NO softkey and adjust dose
below Guardrails Soft Alert .
The bolus dose has been set to a value which is under a Guardrails Soft Alert . Check
the bolus setting, to continue with the bolus confirm OVERRIDE LIMIT by pressing the
YES softkey . If OVERRIDE LIMIT is not required press the NO softkey and adjust dose
above Guardrails Soft Alert .
The bolus dose has been set above a Hard Limit . Check bolus setting and adjust to
appropriate required dose .
The drug concentration has been set above or below a Hard Limit . Check the amount
and total volume and adjust to give the appropriate required concentration .
The patient weight has been set to a value which exceeds or is under a Guardrails Soft
Alert . Check the weight setting, to continue confirm OVERRIDE LIMIT by pressing the
YES softkey . If OVERRIDE LIMIT is not required press the NO softkey and adjust the
value within the limits .
The infusion rate has been set above a Hard Limit . Check infusion setting and adjust to
appropriate required rate .