Alaris Editor Software Profile Drug library

The following drug parameters are only configurable via the Alaris Editor Software, see Alaris Editor Directions For Use for details on how
to configure Profile Drug Library, and are used when the Pump is operated with a drug name selected .
Concentration Units
Concentration Limits
(Min and Max)
Continuous Dose Rate -
Bolus Mode
Bolus Dose -
Bolus Rate -
GH
Occlusion Alarm Pressure
Occlusion Alarm Pressure
Guardrails
*

Profile Syringe Library

The Profile Syringe Library is created from the predefined Master Syringe Library .
Check the boxes of the syringes to be included in the profile . Checking the All Syringes checkbox under Operations selects all the
syringes .
For Syringe brands and sizes available see 'Recognised Syringes' section .
Note: It is recommended that only syringe types and sizes used in the care area are selected .
The unit for concentration parameters
These define the range over which the drug concentration can be modified
during programming of the Pump .
Units
The continuous dose rate units . Can be based on patient weight .
Soft Min*
The continuous dose rate value below which override confirmation is required .
Default
The default continuous dose rate offered when the drug is selected .
Soft Max*
The continuous dose rate value above which override confirmation is required .
Hard Max
The maximum allowed continuous dose rate .
Bolus feature can be set to one of the following options:
Units
The bolus dose units . Can be based on patient weight .
Soft Min*
The bolus dose value below which override confirmation is required .
(Hands-Free only)
Default
The default bolus dose offered .
(Hands-Free only)
Soft Max*
The bolus dose value above which override confirmation is required .
(Hands-Free only)
Hard Max
The maximum allowed bolus dose .
Default
The default value for bolus rate in ml/h .
The default occlusion alarm pressure .
The default occlusion alarm pressure .
BDDF00001 Issue 5
• Disabled
• Hands-On only
• Hands-Free and Hands-On
36/56