Purge

i
The
button allows the delivery of a limited volume of fluid in order to purge the extension set prior to being connected to a patient
or after changing a syringe .
i
1 . Press the
button when the Pump is not infusing . Ensure that the extension set is not connected to the patient .
2 . Press and hold the PURGE softkey until fluid flows and the purging of the extension set is complete . The volume used during
purging will be displayed, but it is not added to the volume infused .
3 . When purging is complete release the PURGE softkey . Press the QUIT softkey to exit back to the main display .
The Pump will not purge if the RATE LOCK has been enabled. During PURGE the pressure limit alarms are temporarily
w
increased to their maximum level.

Volume to be Infused (VTBI)

This option allows a specific volume to be infused to be set . Rate at the end of this VTBI can also be set, selecting from stop, KVO, or
continuous infusion at the set rate .
1 . Press the VTBI softkey to select the volume to be infused option .
2 . Enter the volume to be infused using the
3 . Select the rate at the end of the VTBI using the
4 . Press the OK softkey to confirm and exit the VTBI menu .
Note: When current VTBI has finished, no other infusion will be allowed unless a new VTBI is set or current VTBI is cleared .

Clear Volume

This option enables the volume infused to be cleared . The Dose Infused for a drug is displayed if the Volume Infused is attributable to a
single drug setup . Clearing the volume will display the Dose Infused .
1 . Press the VOLUME softkey to display the CLEAR VOLUME option .
2 . Press the YES softkey to clear the volume . Press the NO softkey to retain the volume .
Note: Selecting YES resets the volume infused in the 24H LOG option .

Rate Lock

If Rate Lock is enabled, when the infusion rate has been set and the infusion started, the rate lock prompt will appear on the display
following any rate titrations or bolus infusions .
To select the rate lock function press the YES softkey to confirm . Press the NO softkey if the rate lock is not required .
When rate lock is enabled, the following are unavailable:
• Changing the infusion rate / titration
• Bolus / purge
• Switching the Pump off
• VTBI over time infusions .
To disable the rate lock if selected:
d
1 . Press the
button to access the options menu .
2 . Select the UNLOCK RATE option using the
To enable the rate lock if not selected:
d
1 . Press the
button to access the options menu .
2 . Select RATE LOCK option using the

Rate Titration

If Rate Titration is enabled the rate can be adjusted while infusing:
1 . Select the new rate using the
2 . The message < START TO CONFIRM > will flash on screen and Pump continues to infuse at the original rate .
b
3 . Press the
button to confirm the new infusion rate and start infusing at the new rate .
Note: Press the QUIT softkey to exit titration and return to original rate .
Note: If the new infusion rate setting exceeds or is under a Guardrails Soft Alert confirmation is required before infusion can start
infusing at the new rate .*
If Rate Titration is disabled the rate can only be adjusted whilst on hold:
h
1 . Press the
button to put the Pump on hold .
2 . Select the new rate using the
b
3 . Press the
button to start infusing at the new rate .
Guardrails
*
f
keys and press the OK softkey .
f
f
keys and press the OK softkey .
f
keys and press the OK softkey .
f
keys .
f
keys .
BDDF00001 Issue 5
keys to scroll through the on-screen choices . The default is stop .
23/56
Alaris™ Syringe Pumps
Basic Features

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Alaris 8003tig03-gAlaris 8003tig03Alaris 8002tig03-g

Table of Contents