Alaris Gh Syringe Pump And Alaris Gh Guardrails Syringe Pump - BD Alaris 8002TIG03 Directions For Use Manual

Syringe pump with plus software mk4
Hide thumbs Also See for Alaris 8002TIG03:
Table of Contents

Advertisement

Alaris GH Syringe Pump and Alaris GH Guardrails Syringe Pump

Start-up Trend . BD Plastipak 5ml @ 0 .1ml/h
0.2
0.1
0.0
0
20
40
Time (mins)
Start-up Trend . BD Plastipak 50 ml @ 1 .0 ml/h
Time (mins)
Start-up Trend . BD Plastipak 50 ml @ 5 .0 ml/h
Time (mins)
60
80
100
120
BDDF00001 Issue 5
Trumpet Curve . BD Plastipak 5ml @ 0 .1ml/h
30.0
20.0
10.0
0.0
-10.0
-20.0
-30.0
0
5
10
15
Observation Window (mins)
Maximum Error
Minimum Error
Trumpet Curve . BD Plastipak 50 ml @ 1 .0 ml/h
Observation Window (mins)
Maximum Error
Minimum Error
Trumpet Curve . BD Plastipak 50 ml @ 5 .0 ml/h
Observation Window (mins)
Maximum Error
Minimum Error
53/56
Alaris™ Syringe Pumps
Trumpet Curves and Start-up Curves
20
25
30
35
Linear Mean = -2 .5%
Linear Mean = -1 .8%
Linear Mean = +0 .2%

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Alaris 8003tig03-gAlaris 8003tig03Alaris 8002tig03-g

Table of Contents