Attaching An External Lens; Anterior Chamber Scan (Chamber View™) - Zeiss CIRRUS HD-OCT 500 User Manual

Hide thumbs Also See for CIRRUS HD-OCT 500:
Table of Contents

Advertisement

6-8
Anterior Chamber Scan (Chamber View™)
2660021169012 Rev. A 2017-12
Anterior Segment Scans
• Wide Angle-to-Angle Scan
Cornea External Lens:
• HD Cornea Scan
• Pachymetry scan

Attaching an External Lens

Anterior Chamber Lens
Lens symbol
English lens
label
To attach an external lens and select a scan
1. Attach the appropriate external lens to the instrument lens mount.
cáÖìêÉ=SJO==^íí~ÅÜáåÖ=~å=bñíÉêáçê=iÉåëI=^åíÉêáçê=`Ü~ãÄÉê=iÉåë=ëÜçïå
2. CIRRUS HD-OCT automatically detects the lens, adjusts lens positioning, and displays
the scans that require the lens in the scan list.Select the scan from the scan list for the
eye you want to scan.
This scan generates a wide field, speckle–reduced raster scan of the anterior chamber at a
depth of 5.8 mm, with higher contrast than the Anterior Segment 5 Line Raster scan. It
uses 20 B-scans, each composed of 1024 A scans, and is 15.5 mm in length when oriented
Cornea Lens
Lens symbol
English lens
label
CIRRUS HD-OCT User Manual