Advanced Rpe Analysis Acceptance Criteria - Zeiss CIRRUS HD-OCT 500 User Manual

Hide thumbs Also See for CIRRUS HD-OCT 500:
Table of Contents

Advertisement

7-6
Criteria for Image Acceptance
OCT fundus image banding examples (A and B) below. As long as there are no saccades,
scans with OCT fundus images like these should be acceptable for analysis and the
operator is advised to save them.
Example A: OCT Fundus image banding
Example B: OCT Fundus image banding

Advanced RPE Analysis Acceptance Criteria

In order to detect sub–RPE illumination, CIRRUS looks for contrast in a slab created below
the RPE. If the retinal tissue is captured too low in the axial field of view of the scan, then
the algorithm will not produce a good result, because there will not be enough sub–RPE
pixels to create good contrast. See
Figure 7-4
below for an example of a scan with the
retina too low for acceptable detection of sub–RPE illumination. If you obtain an image
like this, it is advisable to retake the scan before running Advanced RPE Analysis (see
"Advanced RPE Analysis" on page
8-17).
cáÖìêÉ=TJQ==bñ~ãéäÉ=çÑ=pÅ~å=ïáíÜ=oÉíáå~=qçç=içï=áå=cáÉäÇ
2660021169012 Rev. A 2017-12
CIRRUS HD-OCT User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cirrus hd-oct 5000

Table of Contents