Electrical, Physical and Environmental
Weight
Dimensions
Fixation
Internal fixation focus adjustment
Input devices
Electrical rating (115V)
Fuse rating
Electrical rating (230V)
Convenience Receptacle output ratings
Temperature (transport and storage)
Relative humidity (transport and storage)
Atmospheric pressure (transport and storage)
Temperature (operation)
Relative humidity (operation)
Altitude (operation)
Atmospheric Pressure (operation)
Computer
Room Lighting
Maintenance
Measurement Units
CIRRUS HD-OCT User Manual
Electrical, Physical and Environmental
Model 500
34 kg (76 lbs)
Single Phase, 100/120V~ systems:50–60Hz, 5A
Single Phase, 220/240V~ systems:50–60 Hz, 2.5A
10% to 100%, including condensation
30% to 75%, excluding condensation
• i7 Intel® processor
• Internal storage: > 80,000 scans
• USB ports, 6
• Integrated 19" color flat panel display
• Windows 7 Ultimate, 64 bit
Standard indoor office fluorescent lamp
environment. Not to be used in direct sunlight
Zeiss recommends regular preventative maintenance.
NOTE: Only trained CZMI personnel may perform calibration.
All units on the CIRRUS HD-OCT are measured in the SI format. Unless otherwise noted,
measurements are made in micrometers (μm).
Model 5000
36 kg (80 lbs)
65L x 46W x 53H (cm)
Internal, external
–20D to +20D (diopters)
Keyboard, mouse
Line and Neutral are fused.
N/A
Line and Neutral are fused.
N/A
–40º to +70º C
500 hPa to 1060 hPa
+10º to +35º C
Up to 3000 m above sea level.
700 hPa to 1060 hPa
(e.g., near a window).
2660021169012 Rev. A 2017-12