Forum/Dicom Mode; Transferring Images In Opt Iod And Op Iod Formats - Zeiss CIRRUS HD-OCT 500 User Manual

Hide thumbs Also See for CIRRUS HD-OCT 500:
Table of Contents

Advertisement

FORUM/DICOM Mode

CIRRUS HD-OCT User Manual
You may export repeatedly to the same export folder (on a network drive or the same
removable media), in which case CIRRUS adds only those exams that were not already
exported and does not overwrite any data previously exported. You may interrupt an export
operation and then start it again later, and only the exams not already present in the
export database are exported.
It is possible to export data:
• To a hard-drive on the CIRRUS instrument itself. Do not use the C: drive however.
Specify a location for data export that is not the same location where the local
database is stored.
• Directly to another CIRRUS HD-OCT instrument (see
page
11-5).

Transferring Images in OPT IOD and OP IOD Formats

This function allows the export of image files from the instrument or CIRRUS Review
Software, in a standard DICOM format for viewing on a remote station.
NOTE: This option is not available for the Optic Disc Cube scan and both Anterior Segment
scans.
Terms
• IOD
– Information Object Definition
• OP
– Ophthalmic Photography
• OPT
– Ophthalmic Tomography
• OPT IOD – DICOM standard format for archiving and transferring OCT images as black
and white images
• OP IOD – DICOM standard 8–bit Image format for archiving and transferring the
fundus images as black and white images
Enabling this Operation and Setting Transfer Times
To start using this functionality:
1. Go to Records > Preferences > DICOM Archive and select Send OP and OPT IODs
During Archive (Except After Saving) (see
Importing and Exporting Data
"FORUM/DICOM Mode" on
Figure 11-4
below).
2660021169012 Rev. A 2017-12
11-5

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cirrus hd-oct 5000

Table of Contents