Cirrus Oct Angiography Presets - Zeiss CIRRUS HD-OCT 500 User Manual

Hide thumbs Also See for CIRRUS HD-OCT 500:
Table of Contents

Advertisement

9-4
2660021169012 Rev. A 2017-12
Overview

CIRRUS OCT Angiography Presets

The CIRRUS HD-OCT Angiography and Montage Angio analysis screen includes 12
pre-defined angiography presets, which are the same. The CIRRUS HD-OCT ONH
Angiography includes 4 predefined angiography presets. These slabs are discussed below.
Larger images of each, and a further discussion of the algorithms used to define their
boundaries, can be found in Table 9-4.
• Retina Depth Encoded: This is a color encoded slab with different colors representing
different layers (Red: Superficial; Green: Deep; Blue: Avasculature).
NOTE: This preset is not available for ONH Angiography.
• Retina: This is intended to illustrate vasculature of the entire retina. The inner
boundary is the ILM. The outer boundary is offset above the RPE by 70 μm to minimize
the contribution of the hyper-reflective RPE.
• VRI: This preset is designed to highlight neovascularization above the vitreoretinal
interface (VRI), such as in proliferative diabetic retinopathy. The outer boundary is the
ILM, and the inner boundary is offset above the ILM by 300 μm.
• Superficial: Superficial retinal layer slab. The inner surface is the ILM layer segmented
in the same manner as CIRRUS HD-OCT structural images, such as Macular Cube
512x128. The outer surface is an approximation of the inner plexiform layer (IPL),
which is estimated by the following equation:
Z
=Z
+70%*(T
IPL
ILM
ILM-OPL
Where Z
is the boundary location of the estimated IPL, Z
IPL
location of the ILM, and T
layer (OPL), which is estimated as discussed below.
NOTE: This preset is not available for ONH Angiography.
• Deep: Deep retina layer slab. The inner surface is IPL as described above. The outer
surface is OPL, which is approximated as:
Z
=Z
– 110 μm
OPL
RPEfit
Where Z
is the boundary location of the estimated OPL, and Z
OPL
location of the RPE segmented in the same manner as CIRRUS HD-OCT structural
images.
NOTE: This preset is not available for ONH Angiography.
• Avascular: The inner surface of the Avascular slab is the estimated OPL. The outer
surface is the boundary between the inner and outer segment junctions (IS/OS), the
position of which is estimated as:
Z
= Z
– 70 μm
IS/OS
RPEfit
)
is the thickness between ILM and the outer plexiform
ILM-OPL
is the boundary
ILM
is the boundary
RPEfit
CIRRUS HD-OCT User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cirrus hd-oct 5000

Table of Contents