Width, Height, and A–Scan Adjustments
You can manually adjust the red (Width), blue (Height), and green (A–Scan) sphere
positions by clicking the up/down arrows, or by typing the desired position in the Width,
Height, and A–Scan fields. These fields may provide finer control of the sphere positions.

Mouse Rotates/Mouse Translates

Two buttons allow the user to set the left mouse to either rotate or translate the image on
the screen.

Zoom

Use the slider to zoom in or out on the image.

Straighten Volume Data

reset buttons
cáÖìêÉ=UJQV=sçäìãÉ=píê~áÖÜíÉå=aá~äçÖ
This function allows you to adjust the surface of the scan to be horizontal in those cases
where the retina is highly tilted. Click Auto Straighten to automatically correct the image.
The slider controls, number field, and up/down arrows are used to manually correct the
image by adjusting the X and Z axes. Click Reset to Zero to undo your corrections. The reset
buttons (see figure above) located above each field allow you to reset X and Z axes
separately.
Figure 8-50
illustrates before and after Auto Straighten has been applied. The
left figure shows a tilted image, and the right figure shows the straightened image.