Section 2: Setup meter communications
January 2018
2.1.3
Collect As Found logs and reports
Use the Daniel MeterLink
historical record of the meter a t several velocities within the operating range of the meter to
establish a baseline to be used for the trending of the meter diagnostics and save the following
logs and reports for the "As Found" settings (factory default).
•
•
•
8
Logs/Reports
Maintenance log
System log
Save your meter's configuration file
–
Use the
Tools|Edit/Compare Configuration
View All or Metrology
The
displays the meter's extended configuration parameters. Grayed out
All View
values are read-only, but help describe the configuration. The
displays the path dimensions, pipe diameter, transducer delay parameters, zero flow
calibration coefficients, and flow calibration coefficients.
Click the question mark icon for the data point to display additional information.
–
Click
to save the configuration file. By default, the file is saved to the Data Folder
Save
setup in
File|Program Settings
3410 Series Gas Ultrasonic Meters Operations Manual
menu and collect and save the logs and reports for a
menu to display the dialog. Enable the
radio button and click
Read
.
3-9000-777 Rev E
.
Metrology View
Collect As Found logs and reports