Wet Calibration Piece-Wise Linearization; Figure 5-2 Calibration - Meter Factor: None - Emerson Daniel 3410 Series Operation Manual

Gas ultrasonic flow meters
Hide thumbs Also See for Daniel 3410 Series:
Table of Contents

Advertisement

Section 5: Calibrate and configure the meter
January 2018
2.

Figure 5-2 Calibration - Meter Factor: None

5.1.3

Wet calibration piece-wise linearization

If the 12-point piece-wise linearization (PWL) wet calibration method is selected, then the dry-
calibration gas flow velocity is calculated as shown in
The inputs to the 12-point piece-wise linearization are the (up to) 12 pairs of volumetric flow
rates and meter factors for each flow direction ([
[
FwdFlwRt12
[
RevFlwRt12
points using Daniel MeterLink.
The input flow rates should be entered in descending order (highest first) with their
corresponding meter factors. If these are entered in the wrong order, the meter will still
work but the accuracy may be affected. The meter will not accept negative values, so only
positive values should be entered for both forward and reverse flow rates. If fewer than 12
points are to be used, the unused flow rate and meter factor pairs should be set to 0 and 1
respectively.
44
Select the
Calibration|Meter Factors
,
] for forward flow; [
FwdMtrFctr12
,
] for reverse flow). Refer to
RevMtrFctr12
3410 Series Gas Ultrasonic Meters Operations Manual
menu.
Equation
5-3.
,
FwdFlwRt1
FwdMtrFctr1
,
RevFlwRt1
RevMtrFctr1
Section 5.4
Wet calibration piece-wise linearization
3-9000-777 Rev E
], ...,
], ...,
to configuring these data

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Daniel 3414Daniel 3412Daniel 3411

Table of Contents