6–2–39 Front Bumper Set (FRU No. 310)
Time Required
10 Minutes
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6–62.
1.
Turn OFF the system.
2.
Remove the Front Base Cover (FRU 305). Refer to 6–2–35 on page 6–131.
Unscrew three hexagonal bolts from each Caster Bumper ( 1 – 3 ).
3.
4.
Remove the Front Bumpers.
Reuse the hexagonal bolts removed to install the new front bumper(s).
Tall version console is used the BUMPER LONG SET ASSY.
FRONT BUMPER SET DISASSEMBLY