Crt Cable Set (Fru No. 116) - GE LOGIQ 400 Service Manual

Hide thumbs Also See for LOGIQ 400:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 9
6–2–11 CRT Cable Set (FRU No. 116)
Time Required
0.5 Hours
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6–27.
1.
Turn OFF the system.
2.
Remove the Monitor Assy (FRU 100). Refer to 6–2–1 on page 6–70.
3.
Remove the Monitor Cover Set (FRU 101). Refer to 6–2–2 on page 6–72.
Disconnect three connectors ( 1 – 3 ) from the back side of the CRT Assy.
4.
CRT ASSY DISASSEMBLY
ILLUSTRATION 6–27
6–88
LOGIQ 400 SERVICE MANUAL
2127661