Crt Lamp Assy (Fru No. 113), Crt Lamp (Fru No. 114) - GE LOGIQ 400 Service Manual

Hide thumbs Also See for LOGIQ 400:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 9
6–2–9 CRT Lamp Assy (FRU No. 113), CRT Lamp (FRU No. 114)
Time Required
0.5 Hours
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–25.
1.
Turn OFF the system.
2.
Remove the Monitor Cover L Assy (FRU 104) and Monitor Cover R Assy (FRU 105). Refer to 6–2–2 on page
6–72.
3.
Turn the CRT lamp counterclockwise and remove it.
Unscrew two screws ( 1 and 2 ).
4.
Disconnect one connector ( 3 ).
5.
6.
Remove the CRT Lamp Assy.
CRT LAMP ASSY DISASSEMBLY
ILLUSTRATION 6–25
6–84
LOGIQ 400 SERVICE MANUAL
2127661