Calibration Procedures; Equipment Required - Keysight Technologies 53147A Service Manual

Microwave frequency counter, power meter, dvm
Table of Contents

Advertisement

Calibration Procedures

There are four calibration procedures that can be performed outside of an
Keysight authorized service facility. The following procedures describe how to
calibrate the frequency of the instrument's timebase, the frequency and output
level of the power reference oscillator, and the DVM.

Equipment Required

Table 2-1
calibration procedures in this chapter.
Table 2-1
Recommended Test Equipment and Accessories
Instrument Type
Digital Multimeter
Power Meter
Power Meter
Frequency Counter
Thermistor Mount
Power Sensor
Power Sensor Cable
Range Calibrator
Serial Cable
Precision DC Power Supply + 50 V / – 50 V
Cable
Adapter
[a] Equivalent equipment can be substituted for all instruments and accessories.
[b] One supplied with Keysight 53147A/148A/149A
Keysight 53147A/148A/149A Assembly Level Service Guide
lists the test equipment and accessories needed to perform the
Required Characteristics
6½ digit AC/DC
Range 1 mW
100 kHz to 110 GHz, –70 dBm to +44 dBm
10 Hz to 125 MHz
SWR 1.05 at 50 MHz
Accuracy ±0.5% at 50 MHz
10 MHz to 18 GHz, -30 dBm to +20 dBm
5 ft.
3, 10, 30, 100, and 300 µW
1, 3, 10, 30, and 100 mW
RS-232, DB-25 to RJ-45
BNC (m) to BNC (m)
Type N (m) to BNC (f)
[a]
Model Recommended
Keysight 34401A
Keysight 432A
Keysight E4418B
Keysight 53150A/51A/52A
Keysight 478A-H75 or
478A-H76
Keysight 8481A
Keysight 11730A
Keysight 11683A
Keysight 53150-60215
Keysight 6613C or 6614C
Keysight 10503 series
Keysight E9635A
Service
2
Quantity Required
1
1
1
1
1
1
[b]
1
1
1
1
1
1
59