Performance Test Record; Performance Test Record (Page 3 Of 3) - Keysight Technologies 53147A Service Manual

Microwave frequency counter, power meter, dvm
Table of Contents

Advertisement

Performance Test Record (Page 3 of 3)

Test
Number
Performance Tests
Power Measurement:
3
Power Reference Frequency:
4
Power Reference Level:
5
1.0 mW (0.00 dBm)
Keysight 53147A/148A/149A Assembly Level Service Guide
Complete
3 µW
10 µW
30 µW
100 µW
300 µW
1 mW
3 mW
10 mW
30 mW
100 mW
20.00 dBm
–10.00 dBm
50.0 MHz
Performance Tests - Verifying Specifications
Specifications
± 0.02 dB or ± 0.5%
"
"
"
"
"
"
"
"
"
±0.04 dBm
±0.04 dBm
50.0 MHz ± 1%
(49.5 MHz to 50.5 MHz)
1.0 mW ± 1.2%
0.988 mW to 1.012 mW
1
Test Resul ts
Pass
Fail
_____________________
_____________________
_____________________
_____________________
_____________________
_____________________
_____________________
_____________________
_____________________
_____________________
_____________________
_____________________
_____________________
_____________________