Troubleshooting - Atmos S 201 Thorax Operating Instructions Manual

Hide thumbs Also See for S 201 Thorax:
Table of Contents

Advertisement

10.0 Troubleshooting

Description
Device cannot be switched
on.
Battery does not recharge
and charging symbol does
not light up despite proper
connection to the power
cable.
„Vacuum too low"
„Vacuum too high"
„Battery low"
„Service required"
System shut down.
High temperature of the
device
Leakage test failed.
Flow readout is always in 0
l/min.
40
Possible causes
Battery is completely empty.
Fuse is defective.
Mains cable defective or not
connected properly.
Recharging unit, charging elec-
tronics or battery are internally
defective.
Leakage
Blockage in the hose system.
Hose is kinked.
Bacterial filter at the measuring
channel / filter in the secretion
canister blocked.
pump
Liquids sucked into
.
Excessively high vacuum ap-
plied from the outside.
Ventilation valve is defect.
Battery almost empty.
Watch Dog Jumper not set.
Battery empty.
Ventilation slots are covered.
Internal error
Secretion canister is leaking.
Hose system is not completely
closed.
Component error
Secretion has entered the
device.
Remedy
Connect the power cable to the device
to recharge the battery, note that the
progress can be checked in the upper
left hand corner of the display.
Check the fuses.
Check the mains cable.
Contact the ATMOS service or a certified
service partner. The device must be
checked.
Check all the hoses and the secretion
canister for leakages.
Remove the blockage if necessary. Re-
move the measuring and rinsing channel
from the connection to the secretion
canister. If the filter in the measuring and
rinsing channel is blocked exchange the
hose system.
Check
the bacterial filter at the measuring
channel and the bacterial filter in the secre-
If the bacterial filter at the
tion canister.
measuring channel is blocked change the
hose system. If the bacterial filter in the
secretion canister is blocked change the
secretion canister.
Contact the ATMOS service or a certified
service partner. The device must be
checked.
Check for correct hose connections.
Contact the ATMOS service or a certified
service partner. The device must be
checked.
Connect device to the supply network.
The battery is charged and the soc (state
of charge) is indicated in the display.
Contact the ATMOS service or a certified
service partner. The device must be
checked.
Connect device to the supply network.
The battery is charged and the soc (state
of charge) is indicated in the display.
Please ensure sufficient air ventilation.
Contact the ATMOS service or a certified
service partner. The device must be
checked.
Contact the ATMOS service or a certified
service partner. The device must be
checked.
1) Check whether the flow is also 0 l/min
when the system is open.
2) Contact the ATMOS service or a certi-
fied service partner. The device must be
checked.