Advertisement

4.5 Function

4.5.1 Target vacuum

• Please note, an adjusted target vacuum over -50 mbar may cause pain and injuries to the
patient.
• On the main screen the target vacuum can be set directly by pressing the
• ATTENTION: The change of the target vacuum is immediately effective. There is no confir-
mation necessary.
• The target vacuum can be freely selected between -5 and -100 mbar in steps of 1 mbar.
• If the buttons
and
• The default value is -20 mbar vacuum at delivery. The standard vacuum value can be individ-
ually set in the user menu.

4.5.2 Suction

• When the device is switched on the pump is not activated. This is visible due to the
in the right upper corner. By pressing the
• The symbol
in the right upper corner of the main screen shows that the pump is running.
• By pressing the
button the pump will be stopped.
• The ATMOS
S 201 Thorax has a vacuum controller. This means, that the integrated pump only
®
starts, if the actual vacuum doesn't correspond to the target vacuum. On the other hand the
pumps performance depends on the difference between the actual vacuum and the target
vacuum.
• The vacuum is measured at the patient side of the hose system.
• The bacterial filter is a safe protection against liquid or bacteria penetrating the pump.
are pressed permanently, the increase / decrease will be accelerated.
If target vacuum is adjusted over
-50 mbar the notice appears ,,High
target vacuum is set".
button the pump starts.
and
buttons.
symbol