Display
WARNING
Faulty battery
The device can not be operated!
Battery must be replaced by the service!
WARNING
Device temperature too high
Provide sufficient ventilation!
Check fan!
Cause of error
If the therapy has not been
started after initial operation
the warning "Inactive therapy"
appears.
• Battery is defective. The
device can no longer be used.
Device temperature too high.
• Device is in the sun or near a
heater.
• Ventilation slots are covered.
• Fan is defective
An inspection according to the manufacturer's specifications is
required every year.
This will be displayed on the device.
Low battery capacity appears in the display.
Battery must be replaced by the service.
If target vacuum is adjusted over -50 mbar the notice appears ,,High
target vacuum is set".
Troubleshooting
Start therapy by pressing the
button.
• Please contact the ATMOS
service.
• Place the device in a cooler
location.
• Please ensure sufficient air
ventilation.
• Please contact the ATMOS
service.