6.
Optional accessories

6.4 Section waste tray

53
18

6.5 Backlighting

1
2
3
40
39
Fig. 37
1
5
4
Fig. 38
• Push the section waste tray (18) from the
front to the microtome base plate (53) until it
is held in place by the two magnets (39) (on
the front of the microtome base plate).
To remove the section waste tray, lift it
slightly and pull it off towards the back.
The backlighting system is inserted
into the one-piece knifeholder base
from the front.
• Remove the two screws (1) using a slotted
screwdriver and then remove the cover
plate (2).
• Insert backlighting (3) into the recess.
• Connect the plug (4) of the backlighting sys-
tem into the socket (5) on the microtome.
Plug the power supply unit into the mains
socket.
The backlighting illuminates once the micro-
tome is turned on with the mains switch.
NEVER use the backlighting device
in conjunction with the SuperMega
cassette clamp.