5.1.5 Specimen holder with precision orientation

In the quick clamping device of the specimen holder fixture with precision orientation, all
object clamps available as optional accessories can be used (implemented).
32
29
Display of the zero position
For better display of the zero position, the orien-
tation has two red indicators (32).
When both indicators are visible and both set-
screws are in zero position at the same time
(notch point, white marking on " "), the speci-
men is in zero position.
When the large standard clamp
(50 x 55 mm) is used, the specimen
orientation of 8° in north-south
direction is no longer possible.
The usable angle is only about 4° in
this case.
Leica RM2235
30
32
31
Fig. 15
The object orientation allows for simple position
correction of the specimen surface when the
specimen is clamped into place.
Orienting the specimen
Specimen blocks must not be ori-
ented during the retraction phase!
If a block is oriented during retrac-
tion, the block will advance by the
retraction value PLUS the selected
section thickness before the next
section.
This may cause damage to both
specimen and knife!
• Raise the object head to the upper end posi-
tion and activate the handwheel lock.
• To release the clamp, turn the eccentric le-
ver (29) forwards.
• Turn setscrew (30) to orient the specimen in
north-south direction. Turn setscrew (31) to
orient the specimen in east-west direction.
Each complete turn of the screw inclines the
specimen by 2°. A total of 4 complete turns =
8° are possible in every direction. The accu-
racy is approximately ± 0.5°.
For ease of estimation, there is a white
marking on the handle and a notch point that
is noticeable during turning.
• To lock the current orientation, turn the ec-
centric lever (29) backwards.