Maintenance Instructions - Leica RM2235 Instruction Manual

Rotary microtome
Hide thumbs Also See for RM2235:
Table of Contents

Advertisement

Universal cassette clamp
60
13
Fig. 43

8.2 Maintenance instructions

Only authorized and qualified service personnel may access the internal components of
the instrument for service and repair!
The instrument is basically maintenance-free.
To ensure trouble-free operation of the instrument over a long period of time, the following is recom-
mended by Leica:
1. Thoroughly clean the instrument on a daily
basis.
2. Enter into a service contract at the latest at
the end of the warranty period. For more in-
formation, please contact your local Leica
technical service center.
We recommend the following inspection intervals depending on the general workload of the instru-
ment as defined in the two categories listed below:
Number of sections per day:
Workload (hours per day):
Sectioning speed:
Specimen material:
Maintenance:
Leica RM2235
• Detach cassette clamp for a thorough cleaning, removing all paraffin
residues (13).
• For cleaning, do not use xylene. Use xylene substitutes or paraffin
removers such as "Para Gard."
• The cassette clamp (13) can also be placed in an oven heated to a
maximum of 65 °C, until the liquid wax escapes.
• Remove paraffin residues with a dry cloth.
• After such a cleaning procedure in an oven, always be sure to lubri-
cate the axle and spring of the clamping lever (60) (see also
8.3).
Category I
> 8.000 sections per day
> 5 hours per day
Predominantly high sectioning speed
Working with soft and hard specimens
Every 12 months
8. Cleaning and maintenance
3. Have the instrument inspected on a regular
basis by a qualified service technician au-
thorized by Leica.
The intervals depend on how heavily the in-
strument is used.
Category II
< 8.000 sections per day
< 5 h/day
Slow to medium sectioning speed.
Working mostly with soft specimens.
Every 24 months