2. Safety

2.3 Integrated safety devices

5
12
10
3
Fig. 2
Locking the handwheel
There are two ways of locking the handwheel
(12):
Using the lever (3) on the right side of the micro-
tome base plate, the handwheel can be braked
in almost any position.
• To brake, rotate the lever in a counterclock-
wise direction to position
Caution!
The braking lever (3) must be exact-
ly in position
wheel brake is applied correctly.
If the lever is moved beyond this
point, it is possible that the hand-
wheel is no longer braked.
• To unlock the handwheel brake turn the
lever (3) back to its original position. Position
.
• To lock the handwheel, press the lever (5)
outwards and continue to turn the hand-
wheel slowly until it locks exactly in the
12 o'clock position.
In either case, the yellow LED in the LOCK field
(4) lights up and the instrument cannot be start-
ed.
If both braking systems are used
simultaneously, always bring the
lever (3) into position
release the brakes. Otherwise it is
possible that the lever (5) cannot be
unlocked.
Instruction manual V 1.1 – 02/2005
.
, so that the hand-
first to

Advertisement