Table 6. Guidance and Manufacturer's Declaration - Electromagnetic Immunity

PLEASE NOTE: The USBCam2 is intended for use in the electromagnetic environment specified
below. The customer or user of the USBCam2 must ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Electrostatic
±6 kV contact
discharge (ESD)
IEC 61000-4-2
±8 kV air
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 610004-4
±1 kV for
input/output lines
Surge
± 1 kV differential
mode
IEC 61000-4-5
± 2kV common
mode
Voltage dips,
< 5% U
short
(>95% dip in
interruptions and
for 0.5
voltage variations
on power supply
< 40% U
input lines
IEC 61000-4-11
(>60% dip in
for 5
< 70% U
(>30% dip in
for 25
< 5% U
(>95% dip in
for 5
Power frequency
3A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80
MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5
GHz
26
Compliance
Level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
± 1 kV differential
mode
± 2kV common
mode
< 5% U
T
U
T)
(>95% dip in
cycle
cycle
for 0.5
< 40% U
T
U
T)
(>60% dip in
cycles
cycles
for 5
< 70% U
T
U
T)
(>30% dip in
cycles
cycles
for 25
< 5% U
T
U
T)
(>95% dip in
sec
sec
for 5
3A/m
3 Vrms
3 V/m
B1051061 Rev. A
Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
T
typical commercial or hospital
U
T)
environment. If the user of the USBCam2
requires continued operation during
mains interruptions, it is recommended
that the PC workstation to which the
T
U
USBCam2 is connected be powered from
T)
an uninterruptible power supply or
battery.
NOTE: U
is the AC mains voltage prior
T
T
to application of the test level.
U
T)
T
U
T)
Portable and mobile RF communication
equipment should be used no closer to
any part of the USBCam2, including its
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance:
d= 1.2 √ P
d= 1.2 √ P for 80 MHz to 800Mhz
d= 2.3 √ P for 800 MHz to 2.5Ghz
Where P is the maximum output rating of
the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation in meters (m).