14-58
SCPI Command Reference

Abort source/measure cycle

ABORt
Description
Program trigger model
COUNt <n>
:TRIGger[:SEQuence[1]]:COUNt <n>
Parameters
NOTE
Query
Description
SOURce <name>
:TRIGger[:SEQuence[1]]:SOURce <name>
Parameters
Abort operation
When this action command is sent, the Model 2520 aborts operation and
returns to the idle state.
A faster way to return to idle is to use the DCL or SDC command.
<n> =
1 to 5000
DEFault
MINimum
MAXimum
INFinite
The product of the trigger count and the number of sweep points
(SOUR1:SWE:POIN) cannot exceed 5000.
:COUNt?
:COUNt? DEFault
:COUNt? MINimum
:COUNt? MAXimum
This command is used to specify how many times the unit cycles
through the trigger layer. For example, if the trigger count is set to 2, the
unit will cycle through the trigger layer twice.
TRIGger:COUNt INFinite can be used for repetitive source waveforms
or for long tests where only the last reading is important.
<name> =
IMMediate
TLINk
TIMer
MANual
BUS
NSTest
PSTest
BSTest
Model 2520 User's Manual
Set trigger count
Specify count
Sets count to 1
Sets count to 1
5000
Run continuously
Queries programmed count
Queries *RST default count
Queries lowest allowable count
Queries largest allowable count
Specify trigger event control source
Pass operation through immediately
Select Trigger Link trigger as event
Select timer as event
Select manual event
Select bus trigger as event
Select low SOT pulse as event
Select high SOT pulse as event
Select high or low SOT pulse as event