Posetup - Keithley 2520 User Manual

Pulsed laser diode test system
Hide thumbs Also See for 2520:
Table of Contents

Advertisement

14-52
SCPI Command Reference

POSetup

:SYSTem:POSetup <name>
Parameters
Query
Description
Error queue
NOTE
[:NEXT]?
:SYSTem:ERRor[:NEXT]?
Description
NOTE
ALL?
:SYSTem:ERRor:ALL?
Description
<name> =
RST
PRESet
SAV0
SAV1
SAV2
SAV3
SAV4
:POSetup?
This command is used to select the power-on defaults. With RST
selected, the instrument powers up to the *RST default conditions. With
PRES selected, the instrument powers up to the :SYStem:PRESet
default conditions. Default conditions are listed in the SCPI tables
(Table 14-1
through
With the SAV0-4 parameters specified, the instrument powers-on to the
setup that is saved in the specified location using the *SAV command.
See
Section 11
for details on the error queue.
As error and status messages occur, they are placed in the Error Queue.
The Error Queue is a first-in, first-out (FIFO) register that can hold up to
10 messages. After sending this command and addressing the Model
2520 to talk, the oldest message is sent to the computer and is then
removed from the queue.
The :STATus:Queue? command performs the same function as
:SYSTem:ERRor[:NEXT]?. See
This query command is similar to the [:NEXT]? command except that
all messages in the Error Queue are sent to the computer when the
Model 2520 is addressed to talk. All messages are removed from the
queue.
Program power-on defaults
Power-up to *RST defaults
Power-up to :SYSTem:PRESet defaults
Power-up to setup stored at memory location 0
Power-up to setup stored at memory location 1
Power-up to setup stored at memory location 2
Power-up to setup stored at memory location 3
Power-up to setup stored at memory location 4
Query power-on setup
Table
14-10).
Read oldest error (code and message)
"STATus subsystem," page
Read all errors (codes and messages)
Model 2520 User's Manual
14-49.