GUIDANCE TABLES (continued)
Electromagnetic Emissions
Guidance and Manufacturer's Deceleration - Electromagnetic Emissions
Prima ENT/DNT is intended for use in the electromagnetic environment specified below. The customer or
user of the Prima ENT/DNT should ensure that it is used in such an environment.
IEC 60601 Test
Immunity Test
ESD
± 6kv contact
IEC 61000-4-2
± 8kv Air
EFT
± 2kv Mains
IEC 61000-4-4
± 1kv I/Os
± 1kv Differential
Surge
± 2kv Common
IEC 61000-4-5
Voltage dips,
>95% Dip
Dropouts,
for 0.5 Cycle
IEC 61000-4-11
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
Power Frequency
3A/m
50/60Hz Magnetic
Field IEC 61000-4-8
32
6137000-795
Guidance and Manufacturer's Deceleration
All Equipment and Systems
Compliance Level
Level
± 6kv contact
± 8kv Air
± 2kv Mains
± 1kv I/Os
± 1kv Differential
± 2kv Common
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
3A/m
Prima ENT/DNT
Electromagnetic Environment
- Guidance
Floors should be wood, concrete or ceram-
ic tile. If floors are synthetic, the R/H
should be at least 30%.
Main power quality should be that of a
typical commercial or hospital environ-
ment.
Main power quality should be that of a
typical commercial or hospital environ-
ment.
Main power quality should be that of a
typical commercial or hospital Environ-
ment. If the user of the Prima ENT
/DNT requires continued operation during
power mains interruption, it is recommen-
ded that the Prima ENT/DNT be powered
from an uninterruptable power supply or
battery.
Power frequency magnetic field should be
that of a typical commercial or hospital
Environment.
Issue 1.5