KROHNE OPTIWAVE 7300 C Handbook page 50

Radar level meter for distance, level and volume measurement of liquids, pastes and solids
Hide thumbs Also See for OPTIWAVE 7300 C:
Table of Contents

Advertisement

Advanced setup functions
Function (Fct.)
C.0.0.0
Advanced setup
C.1.0.0
Installation setup
C.1.1.0
Installation type
C.1.2.0
Tank height or
measuring range
C.1.3.0
Application type
C.1.4.0
Stillwell height
C.1.5.0
Stillwell diameter
C.1.6.0
Antenna extension
C.1.7.0
Distance piece
C.1.8.0
Overfill detection
C.1.9.0
Blocking distance
Input Range
Metal tank, plastic tank or
free air application. Metal
tank
Tank height: 0.2 m or 8" to
40 m or 131 ft. 10 m or 33
ft Measuring range: 0.5 m
or 1½ ft to 40 m or 131 ft.
10 m or 33 ft
Process, storage, agitator,
storage + stillwell, process
+ stillwell or agitator +
stillwell. Process
No disturbance, medium
disturbance or heavy
disturbance. No
disturbance
0 to 40 m or 131 ft. 10 m
or 33 ft
20 mm or 0.8" to 1000 mm
or 40". 100 mm or 4"
Blocking distance to 4 m
or 13 ft (single and double
rod), 6 m or 20 ft (coaxial),
35 m or 115 ft (single or
double cable). 3 m or 10
ft
0 mm or 0 ft to 5000 mm
or 16½ ft. 0 mm or 0 ft
Yes or no. Yes
Antenna extension
(C.1.6.0) to Tank height
(C.1.2.0) or 5000 mm or
16½ ft. 0 mm or 0 ft
OPTIWAVE 7300 C
Handbook OPTIWAVE 7300 C
Description
This menu optimizes the configuration of
the device.
This includes parameters for installation
and application conditions. Also for
defining conversion operations.
Defines where the tank is installed. Pastic
tank also refers to tanks that are made
from glass and fibreglass.
Distance from the tank connecting flange
face/thread stop down to the tank bottom.
Application conditions. "Storage" refers to
a flat product surface. "Process" refers to
a disturbed surface without vortex.
"Agitator" refers to a disturbed surface with
vortex and foam.
If free air selected in Installation type
(C.1.1.0), then alternative menu appears.
"No disturbance" means a flat surface.
"Medium" refers to a disturbed surface
without vortex. "Heavy" means a bubbling
surface with vortex and possibly foam.
Defines the height from the tank
connecting flange face/thread stop down
to the end of the mounting tube.
Defines the inner diameter of the tube.
Probe length is the distance from the
flange face/thread stop of the device down
to the bottom end of the probe (including
counterweight for cable versions). If probe
length has been modified, type in the new
value here.
Distance from the upper side of the flange
to the bottom face of the converter
connection. If a different length is installed,
type in the new value.
Using this function, the level is tracked
within the blocking distance but the
displayed output remains fixed at the
blocking distance limit. An error icon will
be displayed to warn the user that the tank
is overfilling.
The non-measuring range at the top of the
probe. It depends on the probe type and
the installation.
50

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents