7 Calibration
Calibration
7.2.2 Calibration using standard reference solutions ("CAL rEF.S" func-
tion)
Menu selection
Selecting the standard reference solution
Calibration
34
Use the
and
Press the
key to save the setting.
The device will then return to measurement mode, or it will display an error
message if necessary.
The resulting relative cell constant is briefly shown and can always be viewed
in the "SEt ConF" menu under the parameter "CELL Corr".
This function enables the user to automatically calibrate the relative cell con-
stant with a selection of standard reference solutions.
Start configuration.
Select submenu "SEt ConF".
Call up parameter "CAL".
Use the
and
Press the
key to call up the parameter "rEF.S".
Use the
and
Menu/parameter
a
The specifications relate to 25 °C. The temperature responses of the standard reference solu-
tions are stored in the device and are automatically compensated.
Press the
key to save the setting.
The device will then return to measurement mode.
Immerse the conductivity measuring cell in the standard reference solution.
Press the
key for >2 s.
keys to enter the value of the reference solution.
keys to select the "REF.S" setting option.
keys to select one of the standard reference solutions:
a
Meaning
Setting option
1413 µS
Reference solution 0.01 M KCl
2760 µS
Reference solution 0.02 M KCl
12.88 mS/cm
Reference solution 0.1 M KCl
50 mS/cm
Seawater comparative solution KCl
111.8 mS/cm
Reference solution 1 M KCl