Electrical Connections; Mains Supply - TESTO CU-2 User Manual

Table of Contents

Advertisement

#
error notation
1
Status PND1 not ready
2
PND1 temperature error
3
Dilution error
4
Motor error
5
Pump error
6
Dilution air flow PND1 failure
10
Dilution air flow PND2 failure
11
Temperature ET wrong
DF PND2 failure deviation
12
>10%
13
CPC status not ready
14
Saturator temp out of range
15
Condenser temp out of range
16
Optics temp out of range
17
Inlet flow rate out of range
18
Aerosol flow rate out of range
19
Laser power low
20
Liquid level low
21
Concentration out of range
22
Multiple Error at CPC
23
No CPC communication
8

Electrical Connections

8.1.

Mains Supply

Connect the power cord plug to a grounded power socket. The IEC mains connector 17) on the
rear side of the testo CU-2 case includes the mains switch 19) and the fuse holder 18). The one
phase power cord delivered with the instrument is equipped with a country-specific plug and
protective earth.
Mains supply voltage: 90 ... 240 VAC, 50/60 Hz, max. 140 VA
Fuse type:
description
failure in dilution factor
Rotating disk is blocked
raw gas pump not running
flow failure or flow too low
Temperature of the evaporation tube differs 10°C
or more from setpoint
Orifice pressure is < 10 or > 90 kPa. Apply
sufficient external vacuum.
value detected by CPC > 1.0E4 #/ccm
CPC is not connected, switched off or cable
broken
slow switching fuse 250 V, 5 A, t, 5 x 20 mm
8 Electrical Connections