Descripton Of All Ak Commands; Control Commands - 'S - TESTO CU-2 User Manual

Table of Contents

Advertisement

CODE:
AIKG
AIKO
ANGA
ALEC
ANOR
ATD1
AVD1
AVD2
ATET
AFD2
7.5.

Descripton of All AK Commands

7.5.1.
Control Commands – 'S'
SREM
command:
SMAN
command:
SRES
command:
SPAU
command:
STBY
command:
description:
concentration integral average calculated since the command SINT
concentration integral average calculated since command SINT or last AIKO
read the result of the last ZERO TEST
read the result of the last LEAK TEST
read the currently set standard conditions: standard temperature and
read PND1 temperature
read setpoint and actual value of primary dilution factor DF
read setpoint and actual value of secondary dilution factor DF
read evaporation tube ET temperature
read the secondary dilution air supply flow Q
REMOTE – set system to AK host remote control mode
SREM K0
response:
MANUAL – disable remote control and set system to manual operation mode
SMAN K0
response: SMAN E
RESET – reset all functions and restart system
system status after reset is MANUAL mode and PAUSE state
SRES K0
response: SRES E
PAUSE – set system to pause state: resting state after system start or
RESET
PND1 disk stopped, sample gas pump off, dilution air pumps off, PND1 and
ET heatings off
SPAU K0
response:
STAND BY – set system to stand by state: ready for measurements
PND1 disk stopped, sample gas and dilution air pumps on, PND1 and ET
heating on
STBY K0
response:
in PND2
AS
SREM E
E: error status 0...9
E: error status 0...9
E: error status 0...9
SPAU E
E: error status 0...9
STBY E
E: error status 0...9
7 AK Host Operation
PND1
PND2
29

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents