set PND1 temperature the
ETD1
setpoints out of range are rounded to the next lowest respectively highest
possible temperature
command:
ETD1 K0 TTT
set PND1 dilution factor
range: 15...300 for 10 cavities disk or 150...3000 for 8 cavities disk (nominal
EVD1
values)
setpoints out of range are rounded to the next lowest respectively highest
possible dilution factor
command:
EVD1 K0 DDD
set PND2 dilution factor
EVD2
setpoints out of range are rounded to the next lowest respectively highest
possible temperature
command:
EVD2 K0 DDD
ETET
set evap. tube ET temperature range: 20...400°C
command:
ETET K0 TTT
set PND1 diluter disk type and if secondary dilution PND2 shall be used or
not
EDST
choose 8 or 10 cavities disk. The disk selection but not the PND2 setting will
be stored and used after reset or system restart. If PND2 is disabled, air
supply pump is switched off and DF
command:
EDST K0 CC D response:
range: 20...150°C
response:
ETD1 E
E: error status 0...9
TTT: temperature [°C] range: 20...150°C ECE
R83: 150°C
format: floating point example: '150.0'
range: 15...300 (10 cav.) or 150...3000 (8 cav.)
response:
EVD1 E
E: error status 0...9
DDD: dilution factor DF PND1 [ - ] range:
15...300 or 150...3000
format: floating point example: '20.0'
range: 1...11
response:
EVD2 E
E: error status 0...9
DDD: dilution factor DF PND2 [ - ] range:
1...11
format: floating point example: '10.0'
response:
ETET E
E: error status 0...9
TTT: temperature [°C] range: 20...400°C ECE
R83: 300...400°C
format: floating point example: '300.0'
= 1
PND2
EDST E
E: error status 0...9
CC:
used rotating disc format: integer,
2 digits '10' for 10 cavities or '08' for 8 cavities
disk
D: '0' for PND2/pump off '1' for PND2/pump
on
7 AK Host Operation