718 30G/100G
Users Manual
Specifications
Specifications are based on a one year calibration cycle
and apply for ambient temperature from +18 °C to +28 °C
unless stated otherwise. "Counts" are the number of
increments or decrements of the least significant digit.
Pressure Sensor Input, 718 30G
Range
-12 to 30 psi
(-82.7 to 206.85 kPa)
Maximum nondestructive pressure: 3X top of range (90
psi, 620 kPa, 6.2 bar)
Temperature coefficient: 0.01 % of range per ° C for
temperature ranges -10 ° C to 18 ° C and 28 ° C to
55 ° C
Pressure Sensor Input, 718 100G
Range
-12 to 100 psi
(-82.7 to 689.5 kPa)
Maximum nondestructive pressure: 2X top of range
(200 psi, 1380 kPa, 13.8 bar)
Temperature coefficient: 0.01 % of range per ° C for
temperature ranges -10 ° C to 18 ° C and 28 ° C to
55 ° C
24
Accuracy
±0.05 % of range
Accuracy
±0.05 % of range
Pressure Module Input, 718 30G and 718 100G
Range
Resolution
(determined by Pressure Module)
DC mA Input, 718 30G and 718 100G
Range
Resolution
24 mA
0.001 mA
Overload protection: 125 mA, 250 V fast acting fuse
Temperature coefficient: 0.005 % of range per ° C for
temperature ranges -10 ° C to 18 ° C and 28 ° C to
55 ° C
Loop Supply 718 30G and 718 100G
24 V dc nominal
Accuracy
Accuracy, ±(% of
Reading + Counts)
0.025 + 1