Troubleshooting
User Messages
Table 3.1 Error Message Chart
Error Number
Problem
1 through 23,
Microprocessor or
25 and 26,
mechanical
28 through 30
malfunction
32, 36, 39 and 40,
44 through 60,
63 through 68
23 and 24, 27, 31,
Microprocessor or
33 through 35,
mechanical
37 and 38,
malfunction
41 through 43, 62
69 through 77
Microprocessor
malfunction
78 through 80
Error during closing of
door
"Door open" detected
81
during run
82, 83
Door does not open
84
Motor over
temperature
85 through 87
Rotor chamber over
temperature
(refrigerated models)
90 through 96
Temperature sensor
malfunction
(refrigerated models)
98
Rotor not recognized
99
Rotor recognized but
incorrect
a. If the recommended action does not correct the problem, call Beckman Coulter Field Service.
b. In the United States, call 1-800-742-2345; outside the U.S., contact your local Beckman Coulter office or visit us at www.beckmancoulter.com.
3-2
a
Result
Deceleration to full
stop; run cannot be
restarted
Deceleration to full
stop; run cannot be
restarted
Run cannot start until
error is cleared
Run cannot start
Maximum deceleration
to full stop
Deceleration to full stop
Deceleration to full stop
Maximum deceleration
to full stop
Maximum deceleration
to full stop
Maximum deceleration
to full stop
Recommended Action
After rotor comes to a complete stop, turn
the power off (O), then back on (I) to reset.
After rotor comes to a complete stop, open
and close the centrifuge door, then restart.
Turn the power off (O), then back on (I) to
reset.
1. Remove debris in latch
2. Close lid quickly.
3. Turn the power off (O), then back on (I)
to reset.
After rotor comes to a complete stop, shut
the centrifuge door, turn the power off (O),
then back on (I) to reset.
See Emergency Access, below
1. Check that ambient temperature is
within a the limits shown in
Specifications in
CHAPTER
2. Check air inlets and exhausts for
obstructions.
After the motor has cooled, restart. If the
problem persists, call Beckman Coulter Field
b
Service.
Turn the power off (O), then back on (I) to
reset.
Check that the installed rotor is usable in the
centrifuge. If incorrect, install an authorized
rotor; if rotor is authorized and problem
persists, call Beckman Coulter Field Service.
Input correct parameters for the rotor in use,
then restart.
1.
PN B05529AE