Vega VEGAPULS 62 Operating Instructions Manual page 68

Radar sensor for continuous level measurement of liquids
Hide thumbs Also See for VEGAPULS 62:
Table of Contents

Advertisement

9 Diagnosis, asset management and service
Failure (failure)
Code
Cause
Text message
F013
no measured value
available
F017
Adjustment span
too small
F025
Error in the lineari-
zation table
F036
No operable soft-
ware
F040
Error in the elec-
tronics
F080
F105
Determine meas-
ured value
F113
Communication
error
F125
Impermissible elec-
tronics temperature
F260
Error in the cali-
bration
F261
Error in the config-
uration
68
This status message is inactive by default. It can be activated by the
user via PACTware/DTM or EDD.
Maintenance: Due to external influences, the instrument function
is limited. The measurement is affected, but the measured value is
still valid. Plan in maintenance for the instrument because a failure is
expected in the near future (e.g. due to buildup).
This status message is inactive by default. It can be activated by the
user via PACTware/DTM or EDD.
Sensor does not detect an echo
during operation
Antenna system dirty or defec-
tive
Adjustment not within specifica-
tion
Index markers are not continu-
ously rising, for example illogical
value pairs
Failed or interrupted software
update
Hardware defect
General software error
The instrument is still in the
start phase, the measured value
could not yet be determined
Error in the internal instrument
communication
Temperature of the electronics
in the non-specified range
Error in the calibration carried
out in the factory
Error in the EEPROM
Error during setup
False signal suppression faulty
Error when carrying out a reset
Rectification
Check or correct installation
and/or parameter settings
Clean or exchange process
component or antenna
Change adjustment according
to the limit values (differ-
ence between min. and max.
≥ 10 mm)
Check linearisation table
Delete table/Create new
Repeat software update
Check electronics version
Exchanging the electronics
Send instrument for repair
Exchanging the electronics
Send instrument for repair
Disconnect operating voltage
briefly
Wait for the end of the switch-on
phase
Duration up to approx. 3 min.
depending on the version and
parameter settings
Disconnect operating voltage
briefly
Send instrument for repair
Check ambient temperature
Insulate electronics
Use instrument with higher
temperature range
Exchanging the electronics
Send instrument for repair
Repeat setup
Repeat reset
DevSpec
Diagnosis Bits
Bit 0
Bit 1
Bit 2
Bit 3
Bit 4
Bit 5
Bit 6
Bit 12
Bit 7
Bit 8
Bit 9
VEGAPULS 62 • Profibus PA