3
Specifications and Characteristics

N7024A Probe Specifications and Characteristics

Table 9
N7024A Environmental Specifications
Attribute
Use
Temperature
Altitude
Humidity
Pollution Degree
30
Specification
With N7022A Main Cable / N7021A Pigtail Cable
Operating: –40 °C to +85 °C (main cable/pigtail
cable)
Non-operating: –40 °C to +85 °C (main
cable/pigtail cable)
Operating: 4600 m (15,092 feet)
Non-operating: 15,300 m (50,196 feet)
Operating: 25 – 85% room humidity
Non-operating: 25 – 85% room humidity
For indoor use only
Operating: –10 °C to +55 °C (probe pod)
Non-operating: –40 °C to +70 °C (probe pod)
temperatures up to 31 °C, decreasing linearly to
Pollution Degree 2
N7020A and N7024A Power Rail Probes User's Guide
Specification
With N7023A / N7032A / N7033A Browser
Operating: –10 °C to +55 °C (browser)
Non-operating: –40 °C to +70 °C (browser)
Operating: 4600 m (15,092 feet)
Non-operating: 15,300 m (50,196 feet)
Operating: 80% room humidity for
40% at 50 °C
Non-operating: 95% room humidity for
temperatures up to 40 °C