Keysight Technologies N7020A User Manual page 82

Power rail probes
Table of Contents

Advertisement

8
Performance Verification
4
Connect the other end of the calibration coax cable to the Cal Out connector on
the front panel of the oscilloscope.
Cal Out
Connector
Figure 28
5
Press [Auto Scale] on the front panel.
6
Choose Setup > Acquisition, then select the Enabled checkbox for Averaging and set
the # of averages to 1024.
7
Set timebase to 2ns/div.
8
Choose Setup > Math Functions... or Math > Functions....
9
In the Function dialog box, select the tab of the function f1 to define it.
10
From the Function 1 listbox, select Math and then Differentiate.
82
Use the 3.5mm or 2.92mm SMA female to female adapters to make
N OTE
these connections to Cal Out and input channel of oscilloscope.
Calibration setup with the calibration cable
SMA f-f
Adapter
Calibration
Cable
N7020A and N7024A Power Rail Probes User's Guide