From the Source 1 drop-down listbox, select the input channel of the oscilloscope
to which you connected the calibration cable. This will become the source on
which the f1 function will be performed.
12
Deselect the Low Pass and Align Phase checkbox.
13
Select the tab of the function f2 to define it.
14
From the Function 2 listbox, select FFT.
15
Turn the function display on for f2 by selecting the On checkbox.
16
Set the source for f2 to Function 1 and close the Function dialog box. The
oscilloscope is configured as displayed in the figure below.
Figure 29
17
In the f2 display area, set the following for f2:
N7020A and N7024A Power Rail Probes User's Guide
Oscilloscope's user interface as per the functions defined
stop frequency to 10 GHz.
vertical scale to 3dbm/div.
Performance Verification
8