19
Check the frequency response of the N7024A probe. To do this, perform the
following steps.
Figure 31
N7020A and N7024A Power Rail Probes User's Guide
a Disconnect the calibration coax cable and any adapters that you attached to
the input channel of the oscilloscope.
b Connect the N7024A probe to the input channel and calibration coax cable.
You may need to use the N5442A adapter with the probe in case you are
using a V-series or a Z-series oscilloscope.
SMA f-f
Adapter
Calibration setup with the N7024A probe and calibration cable
c Choose Setup > Probe Configuration....
d Select the tab of the oscilloscope's input channel to which the probe is
connected.
e In the Probe Head block, click Select Head.... Select Generic:User Cable to
configure the oscilloscope to not consider any loss from the N7022A main
cable or any probe browsers. Click OK and close the dialog.
Performance Verification
N5422A
Adapter
N7024A
Probe
Calibration
Cable
8