Keysight Technologies N7020A User Manual page 88

Power rail probes
Table of Contents

Advertisement

8
Performance Verification
2
Connect the calibration coax cable to the N7022A main cable that you want to
verify. Use an SMA f-f adapter to connect the two cables.
3
Then connect this main cable to the N7024A probe.
This hardware setup is illustrated in the figure that follows.
Figure 33
4
Choose Setup > Probe Configuration....
5
Select the tab of the oscilloscope's input channel to which the probe is connected.
6
In the Probe Head block, click Select Head.... Select N7022A:N7024A Main Cable to
configure the oscilloscope to consider the nominal loss from the N7022A main
cable. Click OK and close the dialog.
7
The f2 trace now has the frequency response with the N7024A probe attached to
the N7022A main cable. The already saved and displayed memory 1 trace has the
frequency response of the coax test cable. See
88
SMA f-f
Adapter
Calibration
Cable
Calibration setup with the N7024A probe, main cable, and calibration cable
N7024A
Probe
N7022A Main
Cable
Figure
32.
N7020A and N7024A Power Rail Probes User's Guide