5.17.7 Using the Sip-N-Puff Head Array

WARNING!
Risk of serious injury
Proximity sensors are sensitive to water. If
enough water is present close to sensors, they
may be activated and mobility device may start
moving unintentionally.
– Do not operate Head Array with wet hair.
– Do not operate Head Array in wet weather.
– Do not operate Head Array in any circumstances
where water may come close to sensors.
WARNING!
Risk of serious injury
Sensor pads are made of water resistant vinyl
to get water quickly run off the pads before
activating sensors.
If sensor pads are damaged, water may get in and
mobility device may start driving unintentionally.
If sensor pads are covered by water absorbing
material, mobility device may start driving
unintentionally.
– Do not operate Head Array if sensor pads are
damaged. Change sensor pads immediately.
– Do not cover sensor pads with any material.
88
CAUTION!
Risk of Injury or Damage
Improper mounting or maintenance of the
Sip-N-Puff control including the mouthpiece and
breath tube may cause injury or damage.
Water inside the Sip-N-Puff interface module may
cause damage to the unit.
Excessive saliva residue in the mouthpiece can
reduce performance.
Blockages, a clogged saliva trap or air leaks in
the system may cause Sip-N-Puff not to function
properly.
– Ensure moving parts of the wheelchair,
including the operation of powered seating, DO
NOT pinch or damage the Sip-N-Puff tubing.
– Saliva trap MUST be installed to reduce risk of
water or saliva entering the Sip-N-Puff interface
module.
– Flush the mouthpiece and the breath tube at
least twice a week with warm running water.
Disinfect with oral rinse after cleaning.
– The mouthpiece MUST be completely dry
before installation.
– If Sip-N-Puff does not function properly, inspect
system for blockages, clogged saliva trap or
air leaks. As necessary, replace mouthpiece,
breath tube and saliva trap.
For further maintenance and cleaning instructions,
see 6 Maintenance, page 100 .