WARNING!
Continuous contact between user and cable can
result in frayed cable jacket. This increases risk
of electric shorts.
– Avoid routing the cable where it will come into
continuous contact with the end user.
When installing the bus cable, avoid undue straining of the
cable and connection points. Flexing of the cable should
be minimized wherever possible, to extend service life and
minimize the risk of accidental damage.
Risk of damage to bus cable
Regular bending can damage bus cable
– Use of a cable chain to support the bus cable,
where the cable is subject to regular cyclic
bending, is recommended. The maximum
stretch of the chain should be less than the
length of the bus cable. The force applied to
flex the cable should never exceed 10 N.
Appropriate life testing should be carried out to
determine / confirm the expected service life and
inspection and maintenance schedule.
1637423-E

4.3 Connecting the remote

CAUTION!
Risk of unintended stops
If the plug of the remote cable is broken, the
remote cable may come loose while driving. The
remote could suddenly switch off when losing
power. This forces an unintended stop.
– Always check the plug of the remote for
damage. Contact your provider immediately in
case of a damaged plug.
Risk of damage to the remote
The remote plug and connector socket fit
together in one way only.
– Do not force them together.
1. Lightly push to connect the plug of the remote cable
and the connector socket. The plug must lock in place
with an audible click.