ADInstruments POWERLAB TEACHING SERIES Owner's Manual page 54

Table of Contents

Advertisement

Table 5.1
Er ror
Red Light Sequence
No.
-2
LNG, LNG
-4
LNG, LNG, LNG, LNG
-7
SHRT, SHRT, LNG, LNG, LNG
-8
SHRT, LNG, SHRT, SHRT, SHRT Real–time clock won't reset or increment Non-Fatal
-11
SHRT, LNG, SHRT, LNG, LNG
-12
SHRT, LNG, LNG, SHRT, SHRT
-13
SHRT, LNG, LNG, SHRT, LNG
-14
SHRT, LNG, LNG, LNG, SHRT
-15
SHRT, LNG, LNG, LNG, LNG
-16
LNG, SHRT, SHRT, SHRT, SHRT ADC bad zero: p == value read
-17
LNG, SHRT, SHRT, SHRT, SHRT EEROM wouldn't read
-18
LNG, SHRT, SHRT, LNG, SHRT
-30
LNG, LNG, LNG, LNG, SHRT
50
Meaning
RAM test failed
ROM check sum was wrong
Required resource missing
Command FIFO flags don't read correctly Non-Fatal
ADC FIFO flags don't read correctly
DAC FIFO flags don't read correctly
DAC state machine won't run
ADC state machine won't run
ADC FIFO capacity/speed tests failed
ADC won't read consistently
PowerLab Teaching Series - Owner's Guide
F a t a l /
Error Name
Non-Fatal
Fatal
eRAMFailure
Fatal
eROMCheckBad
Non-Fatal
eRsrcMissing
eClkBad
eCmdFIFO
Non-Fatal
eADCFIFO
Non-Fatal
eDACFIFO
Non-Fatal
eDACSysBad
Non-Fatal
eADCTimeOut
Non-Fatal
eADCZero
Non-Fatal
eEEROMBad
Non-Fatal
eADCSysBad
Non-Fatal
eChanNoisy